Emergency and transport Ventilators

BA2001 MI-EL   BA2001 MA-EL
Function Models CMV + CMVSIG
AUTO Controlled/Assisted Synchronized MV
AMV + PCM + MVM
Oxigen % > 40% if Air Mix= ONe > 90% if Air Mix=OFF
Frequency Adjustable from 5 to 40 Bpm
Flow Adjustable from 2 to 20 lpm
PEEP Adjustable from 0 to 15 mbar
I/E Ratio 1/2 or 1/1
Pressure Limit Adjustable from 5 to 60 mbar
Trigger Adjustable from 0 to -10 mbar
Inhaler Flow n.a.
Alarms Pressuremax(>55mbar for adults
>30mbar for children)
PressureMin(<10 mbar)
Apnoea (spontaneus breathe stopped)
Oxigen source Failure
Battery Low
SNC Hazardous Selection
 
Function Models CMV + CMVSIG
AUTO Controlled/Assisted Synchronized MV
AMV + PCM + MVM
Oxigen % > 40% if Air Mix= ONe > 90% if Air Mix=OFF
Frequency Adjustable from 5 to 40 Bpm
Flow Adjustable from 2 to 20 lpm
PEEP n.a.
I/E Ratio 1/2 or 1/1
Pressure Limit Adjustable from 5 to 60 mbar
Trigger Adjustable from 0 to -10 mbar
Inhaler Flow n.a.
Alarms Pressuremax(>55mbar for adults
>30mbar for children)
PressureMin(<10 mbar)
Apnoea (spontaneus breathe stopped)
Oxigen source Failure
Battery Low
SNC Hazardous Selection

 

BA2001 GA-El   BA2001 RA-EL

Function Models CMV + CMVSIG
AUTO Controlled/Assisted Synchronized MV
AMV + PCM + MVM
CPAP
Oxigen % > 40% if Air Mix= ONe > 90% if Air Mix=OFF
Frequency Adjustable from 5 to 40 Bpm
Flow Adjustable from 2 to 20 lpm
PEEP Adjustable from 0 to 15 mbar
I/E Ratio 1/2 or 1/1
Pressure Limit Adjustable from 5 to 60 mbar
Trigger Adjustable from 0 to -10 mbar
Inhaler Flow Adjustable from 0 to 15 lpm
Alarms Pressuremax(>55mbar for adults
>30mbar for children)
PressureMin(<10 mbar)
Apnoea (spontaneus breathe stopped)
Oxigen source Failure
Battery Low
SNC Hazardous Selection
 
Function Models CMV + CMVSIG
AUTO Controlled/Assisted Synchronized MV
AMV + PCM + MVM
CPAP
Oxigen % > 40% if Air Mix= ONe > 90% if Air Mix=OFF
Frequency Adjustable from 5 to 40 Bpm
Flow Adjustable from 2 to 20 lpm
PEEP Adjustable from 0 to 15 mbar
I/E Ratio 1/2 or 1/1
Pressure Limit Adjustable from 5 to 60 mbar
Trigger Adjustable from 0 to -10 mbar
Inhaler Flow Adjustable from 0 to 15 lpm
Alarms Pressuremax(>55mbar for adults
>30mbar for children)
PressureMin(<10 mbar)
Apnoea (spontaneus breathe stopped)
Oxigen source Failure
Battery Low
SNC Hazardous Selection

 

Spare parts