Inspiratory pattern
|
|
| Inspiratory flow pattern |
square wave
sinusoidal decelerating |
| Inspiratory pressure pattern |
square wave
semi-sinusoidal, linear rise |
Timing
|
0,1...2 sec |
| Inspiratory time |
0,1...2 sec |
| Expiratory time |
0,1...60sec |
| Pressure |
37 x 35 x 44 cm
'plus patient component 9 cm |
| PEEP |
0...30 mbar |
Pmax |
5...60 mbar |
Volumes
|
|
| Tidal volume |
2...150 ml |
Impedence of ventilator
|
|
| Pos. ventilatory-resistance |
0...200 mbar/l/s |
| Pas. ventilatory-elastance |
0...4 mbar/ml |
High-frequency oscillatory ventilation HFOV
|
|
| Oscillation amplitude |
p appr. 90 mbar |
| Frequency |
5...15 Hz |
| Middl. airway pressure (CDP) |
0...35 mbar |
| |
|
Christina
Paediatric Ventilation and Ventilation monitoring
|
 |
|
Technical Description
- Ventilation Modes
Controlled Ventilation CMV / PCV
- Assisted-Controlled Ventilation, S-PCV
- Intermittent-Mandatory Ventilation
- Synchronised-lntermittent-Mandatory Ventilation, S-
IMV
- High-Frequency Ventilation, HFV
- Spontaneous Ventilation, SV/CPAP with Apnoea
Monitoring
- Manual Ventilation with PEEP and Plateau
- Nasal - CPAP and Insufflation
|
Cycle-Controlled Volume
Inspiratory Time 0.10 to 2 seconds
Expiratory Time 0.10 to 59 seconds
Inspiration in %o
Frequency 1 to 300 breaths per minute
Adjustable Flow 0 to 20 l/min
Adjustable Inspiratory Pressure (PLATEAU) 10 to 60
mbar
Adjustable Expiratory Pressure (PEEP) 0 to 15 mbar
Trigger Sensitivity 0.2 to 2 mbar A P
Oxygen-Air Mixer
adjustable from 21 - 100% O2. Automatic switch in the event of gas supply failure with visual and audio alarm
Jet Depression 0 to 6 mbar
for accelerating and supporting expiration
Heat control and Display
adjustable required temperature 30 - 40C with heated humidifier 30 - 39°C,
Humidity Control and Display
from 50 -100% relative humidity relative humidity 95 - 98%,
Patient Component P3, easily replaceable
autoclave up to 134°C, manual respiration with PEEP and Plateau
Heated Patient Hose System P3
for
stabilising temperature and preventing condensate with dual temperature
sensors for humidifier and hose, Safety shut-off at 41 °C, autoclave up
to 134°C
Integrated Ventilation Monitor
10.5
TFT colour monitor for effective representation of all relevant
respiratory parameters such as tidal volume, minute ventilation, flow,
oxygen concentration, temperature, compliance and resistance.
Trend Representation
from peak pressure Pmax and/or mean pressure Pmean and minute volume MV
Alarm and warning function
alarm according to European Norm EN 742, audio alarm according to EN 475
- trigger parameters automatically or manually
adjustable
- peak, mean and minimum pressure
- minute volume
- oxygen concentration
- respiratory gas temperature
- apnea, disconnection
- gas, power and unit failure warning system
Alpha-numeric display
- peak and mean pressure, PEEP, minute
volume, tidal volume
- frequency, I:E ratio or inspiration in %
- oxygen concentration, respiratory gas
temperature
- Resistance, Compliance
Graphs and Respiratory Loops
pressure (P), volume flow (V'} and volume (V), respiratory chart V'-V, P-V or P-V graph modification
- scaling of all axes in three levels
- freeze function, saving and print out,
- measuring function for frozen graphs by
means of measurement and reference
cursor
- numerical display of measured points as well
as their difference and quotients
Data Ports
- analogue ports for pressure and volume flow,
free choice of all monitored parameters
- serial port (RS 232) for pressure, volume
flow and further options
Upper
Monitor shelf of stainless steel, complete unit ready for use, mounting
for undercarriage or wall rails 25 x 10 mm and connection hoses 02/DL
with right angle plugs DIN
|