Galileo Gold
 
Controls
Ventilation modes (S)CMV, SIMV, SPONT, ASV, P-CMV, P-SIMV
APVcmv, APVsimv, DuoPAP. APRV, NIV
Special functions Manual breath, 100% 02,standby, sigh.apnea backup, tube resistance compensation (TRC)
Patient types Adult, pediatric, Infant
(S)CMV and P-CMV rate 5 to 120b/min
SIMV, P-SIMV, and DuoPAP rate 1 to 60 b/min
Tidal volume / target tidal volume (APV modes) 10 to 2000 ml
PEEP/C PAP and Plow
(DuoPAP and APRV)
0 to 50 cmH20
Oxygen 21 to 100%
I:E ratio 1:25 to 4:1
Inspiratory time 0.1 to 10 s (10 to 80% of cycle time)
Pause time 0 to 8 s (0 to 70% of cycle time)
Peak flow 1 to 180 l/min
T low (APRV) 0.2 to 30s
T high (DuoPAP and APRV) 0.1 to 30s
Pressure trigger 0.5 to 10cmH;0 below PEEP/CPAP
Flow trigger 05 to 15 l/min
Automatic base flow 4 to 30 1/min, depending on flow trigger setting
Pressure control 5 to 100 cmH2O, added to PEEP/ CPAP
P high (DuoPAP and APRV) 0 to 50 cmH20
Pressure support 0 to 100 cmH,0, added to PEEP/CPAP
Pressure ramp 25 to 200 rns
Expiratory trigger sensitivity 5 to 70% of inspiratory peak flow
% minute volume (ASV) 10 to 350%
Flow patterns Sine, square, 100 decelerating, 50% decelerating
Monitoring
Pressure Peak, mean, minimum, plateau, PEEP/CPAP, AutoPEEP
Flow Inspiratory peak, expiratory peak
Volume Expiratory tidal volume, expiratory minute volume, leakage volume
Time Inspiratory, expiratory, I:E ratio, total frequency, spontaneous breath frequency
Oxygen Airway oxygen concentration
Lung function parameters Inspiratory resistance, expiratory resistance, static compliance, inspiratory time constant, expiratory time constant, imposed work of breathing, pressure time product, rapid shallow breathing index, P.01
Real-time waveforms/loops Simultaneous display of up to three waveforms or one loop based on: volume, flow, airway pressure, or auxiliary pressure
Yrending Simultaneous display of up to three parameter trends, selected from 26 possible monitoring parameters, for 1, 12, or 24 hours
Others Waveform freeze and cursor function, inspiratory/expjratpry hold
Flow Sensor dead space 2 ml (infant), 9 ml (pediatric/adult)
  Pulmonary function assessment
P/V Tool Automatic maneuver to assess static
compliance
Alarms
Operator-adjustable low/high minute volume, low/high pressure, low/high tidal volume, low/high rate, apnea time, air trapping
Special alarms Oxygen concentration, disconnection, loss of PEEP, exhalation obstruction, check setings, Flow Sensor alarms, ASV/APV alarms, power supply, baterries, oxygen/air supplies
  Event log Storage and display of up to 1000 events with time stamp
  Electrical and gas supplies
  Input voltage   Storage and display of up to 1000 events with time stamp
  Electrical and gas supply
Input voltage 100 to 240 V~ +/-10%, 50/60 Hz
Power consumption 210 VA maximum
Backup battery time 1 hour typical with fully charged battery
Oxygen  and air supply 200 to 1000 kPa (29 to 145psi)
  Environment
Temperature 10 to 40 C (operating), -10 to 60 C (storage)
Humidity 30 to 75 % noncondensing (operating),
5 to 85 % noncondensing (storage)
Altitude Up to 3000 m (9843 ft)
  Physical dimensions IEC 601-1/EN 60601-1, IEC 601-1-2/
EN 60601-1-2, EN 794-1,
ISO 7767, EN 60601-2-12
  W x D x H 440 x 620 x 1540 mm (17.3 x 24.4 x
60.6 in.) with standard trolley
440 x 620 x 141 5 mm (17.3 x 24.4 x
55.7  in.) with short trolley
400 x 290 x 730 mm (1 5.7 x 11.4 x
28. / in.) with shelf mount
  Weight 48 kg (105 lb)with standard trolley,
27 kg (59.5 Ib) with shelf mount
Main patient outlet  
Air and oxygen inlets DISS male, NIST (option)
  Hardware options Nebulizer; communications, interface including RS-232C port,-remote nurse's call, and I:E ratio
  Standards IEC 60601 -I/EN 60601-1 and Amendments Al andA2, IEC 60601-1-2/EN 60601-1-2, IEC 60601-1-4/EN 60601-1-4, IEC 60601-2-12, EN 794-1, C22.2 No. 601.1, UL 2601-1, ISO 7767
Gold-specific features Tubing resistance compensation (TRC), PA/ Tool, noninvasive ventilation (NIV)
Note: 1 hPa = 1 mbar =1 cm H2O
    Raphael Silver
 
Controls
Ventilation modes (S)CMV+, SIMV+, PCV+, PSIMV+, SPONT
Special modes ASV, DuoPAP, APRV
Special functions Manual breath, inspirator hold, nebulizer,
100% 02, stand-by, sigh, apnea backup
(S)CMV+rate 8 to 80 b/min
PCV+ rate 4 to 80 b/min
SIMV+, DuoPAPrate 1 to 80 b/min
Tidal volume 50 to 2000 ml
PEEP/CPAP and Plow Oto 35cmH2O
Oxygen 21 to 100%
I:E ratio 1:9 to 4:1 ((S)CMV+, PCV+)
Inspiratory time 0.1 to3.2s(SIMV+,PSIMV+)
Tlow (APRV) 0.2 to 30 s
Thigh (DuoPAP and APRV) 0.1 to 30s
Trigger Off, 2 to 101/min
Pressure control 5 to 50 cmH,0 above PEEP/CPAP
Pressure support 0 to 50 cmH,0 above PEEP/CPAP or Plow
Phigh (DuoPAP and APRV) 0 to 75 cmH20
Bodyweight 5 to 200 kg
% minute volume (ASV) 10 to 350%
Flow (automatic) Oto 120 l/min typical, 1801/min maximum
Patient monitoring
Real-time curves Volume, flow, pressure
Loops Pressure-volume, volume-flow, pressure-flow
Trends 1 h, 12 h, 24 h, of 16 monitored parameters
ASV graphics Target and actual values for minute ventilation, tidal volume, and rate
Pressures PEEP/CPAP, peak and mean airway pressures
Volume Spontaneous and total expiratory minute volumes, expiratory tidal volume volume
Flow Inspriatory flow
Time 1:E ratio, inspiratory time, total frequency, spontaneous frequency
Others Resistance, compliance, AutoPEEP, expiratory time constant, trigger, 0,%
Alarms
Lower/upper expiratory minute volume 0.1 to 50 l/min
High pressure 15 to 80 cmH20
High rate 0 to 99 b/min
02% ±5% of setting, minimum 18%
Apnea; 20 s
Special Disconnection, Flow Sensor, pressure limitation, gas supply, electrical supply, battery, exhalation obstructed, ASV messages
Event log Storage and display of up to 1000 events with time stamp
Electrical and gas supply  
Input voltage 100 to 125 and 200 to 240 V~, 50/60 Hz
Power consumption 40 VA typical
Backup battery time 60 minutes typical
O2 and air supply 200 to 600 kPa (29toS6pst)
Physical dimensions  
W x D x H 23 x 53 x 35 cm (9.1 x 20.9 x 13.8 in.) 46 x 66 x 140 cm (18.1 x 28.Ox 55.1 in.) on trolley
Weight Standards 17 kg (37 lb)
46 kg (101 Ib) with trolley
Standards IEC 601-1/EN 60601-1, IEC 601-1-2/
EN 60601-1-2, EN 794-1,
ISO 7767, EN 60601-2-12
Hardware options Communications interface (RS 232, nurse's call, I:E ratio)
Silver-specific features ASV, DuoPAP, and APRV modes; loops; trends; ASV graphics

Arabella
Controls
Flow 0 to 10 l/min
Oxygen 21 to 100%
  Monitoring
NCPAP 0 to 12 cmH2O
Oxygen 21 to 100 %
  Alarms  
High pressure 3 cmH20 above set NCPAP level
Low pressure 2 cmHjO below set NCPAP level
Independent zero pressure alarm
High and low Oxygen ±5% of set Oxygen level
Alarm silence 30s
Internal pop-off valve 150 cmH20
Safety limit valve and alarm Terminates flow if pressure exceeds 11 cmH^O
(reinitiates gas flow after 3 s)
Gas supply Pressure difference between air and oxygen
Electronic displays  
Brightness Auto-dimming
Pressure resolution ±1 cm H2O
Oxygen resolution ±1 %
Pneumatic and electrical supplies  
Air/Oxygen supplies 207 to 510 kPa (2 to 5 bar, 30 to 75 psi)
Electrical power 110 to 240 V ac, 50/60 Hz, 22 VA
Physical characteristics Inspriatory flow
Dimensions (W x D x H 214 x 175x 145 mm (8,4 x 6.9 x 5.7 in.)
Weight 7,1 kg (15.6 lb)
Environmrent  
Operating temperature 10 to 40 C at 0 to 75 relH, noncondensing
Storage temperature -10 to 60 C at 0 to 75 relH, noncondensing
Atmospheric pressure 700 to 1060hPa(10to 15 psi)
Standards ±5% of setting, minimum 18%
IEC 60601-17 EN60601-1, IEC 60601-1-2/EN6060M-2
ARABELLA Universal Generator
  Prongs sizes   XL, L, M, S
ARABELLA Delivery Circuit
Voltages 16, 21, and 28 V
Compatibility of 21 V version Fisher & Paykel MR600, MR620, MR630, MR700,
MR720, and MR730; Hudson RCl 380-88
Compatibility of 16 V version Hudson RCl 380-90
Compatibility of 28 V version Marquest SCT2000 and SCT30QO
ARABELLA Caps 
Cap size 1 Lemon yellow
Cap size 2 Blueberry
Cap size 3 Raspberry
Cap size 4 Lime Green
Cap size 5 Strawberry
ARABELLA five leg stand   
Pole mount for ARABELLA Monitoring Gas Mixer
Note. 1 tiPa - 1 mbar - 1 cmH,O     